Salivary Cortisol Cut-Off Accurately Identifies Cushing's Syndrome and
Late-night salivary cortisol measured using a second-generation electrochemiluminescence immunoassay demonstrated high diagnostic accuracy for detecting Cushing’s syndrome at an established cut-off of 0.204 micrograms per deciliter, according to a diagnostic a
Illustrative imageLate-night salivary cortisol measured using a second-generation electrochemiluminescence immunoassay demonstrated high diagnostic accuracy for detecting Cushing’s syndrome at an established cut-off of 0.204 micrograms per deciliter, according to a diagnostic accuracy investigation conducted at a tertiary care center in Southern India.
The primary objectives of the clinical investigation centered on evaluating ECLIA-II against liquid chromatography-tandem mass spectrometry, commonly abbreviated as LC-MS/MS, while deriving and validating a robust threshold for the diagnosis of Cushing’s syndrome. In addition to primary diagnostic screening, the investigators sought to examine whether late-night salivary cortisol concentrations could serve a secondary clinical function by helping clinicians discriminate between differing forms of adrenocorticotropic hormone-dependent Cushing’s syndrome, specifically separating pituitary-driven Cushing’s disease from ectopic presentations.
In the prospective derivation cohort, researchers enrolled 43 patients confirmed to have Cushing’s syndrome, along with two distinct control arms consisting of 43 individuals categorized as obese controls and 42 individuals designated as normal body mass index controls. In this derivation cohort, late-night salivary cortisol concentrations were quantified using both the ECLIA-II immunoassay and LC-MS/MS platforms to directly contrast assay performance, where an initial ECLIA-II diagnostic cut-off of at least 0.135 micrograms per deciliter achieved sensitivity and specificity values of at least 90 percent for the detection of Cushing’s syndrome.
Direct method comparison revealed meaningful discrepancies between the two analytical measurement techniques when applied to salivary samples.
The clinical performance of the immunoassay threshold was subsequently evaluated within an independent retrospective validation cohort comprising 225 patients, which included 112 individuals with confirmed Cushing’s syndrome and 113 control subjects. In this validation cohort, the diagnostic threshold for ECLIA-II was adjusted to 0.204 micrograms per deciliter, producing a verified sensitivity of 96 percent and a specificity of 90 percent for diagnosing Cushing’s syndrome, confirming the diagnostic accuracy of the revised threshold in a larger testing population.
Evaluation of all adrenocorticotropic hormone-dependent Cushing’s syndrome cases pooled across both cohorts demonstrated additional clinical utility in identifying ectopic pathology. Analysis indicated that a late-night salivary cortisol threshold greater than 1.015 micrograms per deciliter predicted ectopic Cushing’s syndrome with 88 percent sensitivity, 82 percent specificity, and a negative predictive value of 97 percent compared to Cushing’s disease, suggesting this specific cut-off may prompt clinicians to initiate early diagnostic work-up for an ectopic source.
The diagnostic accuracy study was conducted by investigators at the Christian Medical College and Hospital in Vellore, India, with research collaboration from the Centre for Stem Cell Research, a unit of inStem Bengaluru.
In their overall conclusions, the investigators emphasized that the validated cut-off of 0.204 micrograms per deciliter provides robust diagnostic accuracy for confirming Cushing’s syndrome in clinical practice. In addition, the findings indicate that late-night salivary cortisol testing offers meaningful discriminative capability when evaluating adrenocorticotropic hormone-dependent cases.
Subject of Research: Diagnostic accuracy and differentiation utility of late-night salivary cortisol in Cushing's syndrome
Article Title: Validation of late-night salivary cortisol for the diagnosis of Cushing’s syndrome and to explore its utility to differentiate between pituitary and ectopic ACTH-dependent Cushing syndrome
Article References: Agarwal, K., Rajan, R., Govindachari, P. T., Jose, A., Kapoor, N., Cherian, K. E., Prabhu, K., Christudoss, P., Jacob, P. M., Hephzibah, J., Jebasingh, F., Thomas, N., & Asha, H. S. (2026). Validation of late-night salivary cortisol for the diagnosis of Cushing’s syndrome and to explore its utility to differentiate between pituitary and ectopic ACTH-dependent Cushing syndrome. Frontiers in Endocrinology, 17, Article 2018241. https://doi.org/10.3389/fendo.2026.2018241
Image Credits: AI Generated
DOI: 10.3389/fendo.2026.2018241
Keywords: Late-night salivary cortisol, Cushing's syndrome, Cushing's disease, Ectopic Cushing's syndrome, Electrochemiluminescence immunoassay, ECLIA-II, Liquid chromatography-tandem mass spectrometry, LC-MS/MS, ACTH-dependent Cushing's syndrome, Diagnostic accuracy, Bland-Altman analysis, Salivary cortisol cut-off, Diagnostic sensitivity, Diagnostic specificity, Negative predictive value, Derivation cohort, Validation cohort, Pituitary endocrinology
News Source: Ophelia Keating. (November 26, 2026). Salivary Cortisol Cut-Off Accurately Identifies Cushing’s Syndrome and Distinguishes Ectopic Disease. Scienmag.